Distortion-Corrected Diffusion MRI Using Rotated-View EPI and Joint Field-Map/Image Estimation with Gaussian Primitives

Wenqi Huang

Abstract

Echo Planar Imaging (EPI) is the standard acquisition technique for diffusion and functional neuroimaging, enabling rapid imaging but suffering from geometric distortions caused by B0 field inhomogeneities. Existing correction methods first reconstruct distorted images using parallel imaging, then estimate the B0 field and correct the distortion in the image domain. In this sequential process, reconstruction artifacts at high acceleration factors and low SNR at high diffusion b-values degrade B0...

Submitted: July 1, 2026Subjects: Engineering; Biomedical Engineering

Description / Details

Echo Planar Imaging (EPI) is the standard acquisition technique for diffusion and functional neuroimaging, enabling rapid imaging but suffering from geometric distortions caused by B0 field inhomogeneities. Existing correction methods first reconstruct distorted images using parallel imaging, then estimate the B0 field and correct the distortion in the image domain. In this sequential process, reconstruction artifacts at high acceleration factors and low SNR at high diffusion b-values degrade B0 estimation and limit the overall correction quality. We propose a physics-informed framework that jointly estimates the B0 field and distortion-free image directly from k-space data, without depending on an intermediate parallel-imaging reconstruction for the correction. The image and the B0 field are each represented as a superposition of Gaussian primitives embedded within an MRI physics forward model. The explicit, continuous parameterization captures both smooth regions and tissue boundaries and supports rotated-view EPI acquisitions without interpolation. The diffusion-weighted image is modeled as real and non-negative, with the image phase absorbed into a per-shot phase factor. Rotated views distribute distortions across multiple phase-encoding orientations, improving point spread function isotropy and providing stronger constraints for B0 estimation. On in vivo brain diffusion EPI, the proposed method attains the closest brain-boundary agreement with a distortion-free structural reference, with the largest improvement over sequential methods at high b-value and high acceleration. Extensive visual comparisons further show improved detail fidelity and noise suppression.


Source: arXiv:2606.31521v1 - http://arxiv.org/abs/2606.31521v1 PDF: https://arxiv.org/pdf/2606.31521v1 Original Link: http://arxiv.org/abs/2606.31521v1

Please sign in to join the discussion.

No comments yet. Be the first to share your thoughts!

Access Paper
View Source PDF
Submission Info
Date:
Jul 1, 2026
Topic:
Biomedical Engineering
Area:
Engineering
Comments:
0
Bookmark